MCC - Measurements and Tests

Measurements and Tests

Measurements and Testing Services:

Measurements and Tests

Measurement and sampling, particle counting, pressure test, waterproofness, conductivity, bioloads.

Measurements and Testing Services:


Laser particle counter, at 0.3µm, 0.5µm, 1.0µm, 5.0µm according to applicable standards.
Particle counting after depressurization of the gas.
Hot wire anemometer with sensitivity from 0 m/s to 2 m/s.
Hot Wire Anemometer with sensitivity from 0.1 m/s to 15 m/s.
Vane Anemometer with sensitivity from 0.4 m/s to 30 m/s.
Temperature sensor (PT 100), response time Platinum to 0.1 ° C. _Sensor capacities 155 RH, accurate within 2%, with a sensitivity of 1% relative humidity. _Central acquisition Fluke Hydra with scanner 20 lanes. _Intelligent standard RTD Hart 5612/1502 with display probe. _Generator temperature Jofra ATC 155/650 of –30 ° to 155 ° C or 50-650 ° C. _Up to 20 thermocouples teflon type T, accuracy + 0.3 ° C.
Laminar air flows stability and rooms aerolics design, detected by smoke generators. Colored smoke can be used if necessary for filming.
Fresh air intake; recycled air and exhaust air are calculated by measuring and determining the rates of each parameter.
Manometers for Pressure measurement in Pascals, mm/H2O or PSI
Visual Smoke Testing to monitor potential turbulence of the flow, the sealing of the doors, windows and rooms in controlled dust SAS.
Artificial contamination of the air of the ultra-clean zones and calculation of time to return to the initial level of contamination.
Test Dioctylphthalate (DOP) or Polyalphaolefin (known under the name of Emery 3004 PAO) with a photometer and a generator of sensitivity to 99.999% for particles of 0, 3µm.
Sound level meter with a sensitivity between 30 and 150 dba.
Meter sensitivity between 0 and 3000 lux.
Radiometer DM - 254 X measurement of radiation ultraviolet short wavelength between 220 and 280 nanometers. (0-19999 µW/cm²).
Analyzer of vibrations between 5 and 10 000 hertz.
Calibration: 159 Hz, 10 cm/sec ² + 5 %.
Sensibility: 10 millivolts/g (8-14 mV beach of calibration)
Tester of resistance ranges between:
0.1 And 199.99 O step by step of 0.01 O
0.1 And 199.99 O step by step of 0.1 O
0.1 And 1999 O step by step of 1 O
Sensibility + 2 % of reading, + 1 digit
Unit standard pressure and contact time for boxes type RODAC on nutrient agar with neutralizer to inhibit a residual disinfectant action. Mesophilic aerobic Flora . Fungal flora
Bio-impactor device standard in flow rate and sampling time, 100 1/min. with use of boxes type RODAC (possibility to use with boxes of 90 mm) _Type P.C.A. for Mesophilic aerobic Flora _Type SABOURAUD for fungal flora
Analysis of the biological load, more commonly known as "Bioburden" on any type of material and instruments.
Dosing in dialysis water and dialysate fluid according to the European Pharmacopoeia. kinetic chromogenic quantitative method (= 0.005 EU/ml sensitivity threshold).
Withdrawals from tubes containing reagents, or by chromatography.
Central acquisition Fluke Hydra with scanner 20 lanes.
Intelligent standard RTD Hart 5612/1502 with display probe.
Generator temperature Jofra ATC 155/650 of –30 ° to 155 ° C or 50-650 ° C.
Up to 20 thermocouples teflon type T, accuracy + 0.3 ° C.
Rotronic humidity transmitter
Haenni ED 522 pressure transmitter
By gas emission or by smoke and confirmation on chemical indicators.
Validation of sterilization (Peracetic acid or other) times and tightness of the bubble (gaskets and walls). Biological indicators on sterile glass slides on which spores of Bacillus Subtilis Var Niger ATTC 6633 are fixed. Controls (absolute or charcoal) filters particle counting. Microbiological sampling.
Technical assistance and training of personnel by MCC on the behavior and work in ultra clean environments.